Care and Maintenance
4-26 Ultrasound System – Common Service Information
Direction 5444964-100 English
 Rev. 5
Generic procedure
The test verifies the isolation of the power line from the chassis. 
The testing Electrical Safety Analyzer is connected from 
accessible metal parts of the case to ground. Measurements 
should be made under the test conditions specified in: 
• Table 4-6 on page 4-20, or 
• Table 4-7 on page 4-21, as applicable. 
Record the highest reading of current.
1.  Connect the Safety analyzer to wall AC wall outlet.
2.  Plug the equipment under test into the receptacle on the 
panel of the Electrical Safety Analyzer.
3.  Connect the Electrical Safety Analyzer to an accessible 
metal surface of the Ultrasound system using the cable 
provided with the Electrical Safety Analyzer.
4.  Select the “Chassis” or “Enclosure Leakage” function on the 
Electrical Safety Analyzer. 
5.  Test opening and closing the ground with the Ultrasound 
system on and off as indicated in Table 4-6 on page 4-20 or 
Table 4-7 on page 4-21 as applicable. 
NOTE:  For more information, refer to the safety analyzer’s user 
manual that will be used to perform the tests.
The maximum allowable limit for chassis source leakage is 
shown in:
• Table 4-6 on page 4-20, or 
• Table 4-7 on page 4-21, as Chassis/Enclosure Leakage.