Ext. 5V Interface: X45, Terminal 3-4, Intrinsically safe ia 5Vis: (Option)
Phoenix DFK-MSTBVA 2,5/4-G-5,08 jack
Ui ≤ 5,5 V
Ii ≤ not important
Pi ≤ not important
Li ≤ negligible
Ci ≤ negligible
Uo ≤ 5,4 V
Io ≤ 420 mA
Po ≤ 1,25 W
Lo ≤ 0,15mH
Co ≤ 50 uF
Recommended max. cable length 2m (6,5ft)
BS Interface: X45, Terminal 1-2, Intrinsically safe ia:
Phoenix DFK-MSTBVA 2,5/4-G-5,08 jack
Ui ≤ 5,5V
Ii ≤ not important
Pi ≤ not important
Li ≤ negligible
Ci ≤ negligible
Uo ≤ 5,5 V
Io ≤ 36 mA
Po ≤ 50 mW
Lo ≤ 25 mH
Co ≤ 50 uF
Recommended max. cable length 2m (6,5ft)