1.3.5
STATION ESTABLISHMENT CHAPTER 3
GEODIMETER SYSTEM 600
This should be done to get full accuracy of the systems
inherent intelligence.
Comp init
Turn: 200->
I 14:04
The instrument is levelled.
Start compensator calibration by
pressing A/M- or ENT-key.
A beep is heard.
Wait for a double beep after approx.
6-8 sec and the display will change
to.....
Turn the instrument 200 gon (180
°
)
and the display will change to.....
…when the instrument is within 1 gon
of a 200 gon rotation.
A beep is heard and the display will
change to…
Wait for a double beep after approx.
6-8 sec. The display will automatically
change to…
…program 0. The apperance of P0
indicates that the instrument is
sufficiently well levelled and that the
compensator is now engaged. It also
means that the electronic level is in the
"fine mode" in which each individual
left or right movement of the cursor
represents 20 .
Calibration of the dual-axis compensator without servo
cc
Note!
☛
Discon-
nect the
compen-
sator by
setting
function
22=0.
P0 14:05
Temp=20.0_
Comp init
Press A/M
Comp init
Wait
A/M