Operating Instructions Centrifuge Micro III
Gerber Instruments AG 5 Version 1.2 (03/2021)
2.4 Safety instructions for centrifugation
Please check that the centrifuge is set up properly (see 4.1)
Please check before commissioning the centrifuge, rotor and accessories for visible
damage (see 4.2)
The rotor must be loaded symmetrically at equal weights (see 6.1 and 7.).
Avoid corrosion on centrifuge and accessories by careful maintenance (see 8).
Open the centrifuge when it is not in use so that moisture can evaporate.
The centrifuge can be stopped at any time. Press the START/STOP button.
2.5 Safety instructions for cover lock
The centrifuge may only be operated with the cover closed. A cover lock blocks the
cover closure during the run and allows you to open only at standstill.
In case of cut-off of electricity the lid may be opened manually (see 10.2).
2.6 Safety instructions for unbalance switch-off
The centrifuge stops automatically in case of strong vibrations. The switch-off
sensitivity is set at the factory.
If an unbalance switch-off took place, the centrifuge is stopped immediately and a
message appears on the display.
Open the cover and remove the causes of the unbalance.
Close the cover and press START/STOP again.