Germfree - Class II, Type A2 Biosafety Cabinet User Manual - v2.1 - 09SEP2021 - RMH
Class II, Type A2 Biosafety Cabinet
User Manual

Questions and Answers

  • J
    James McknightAug 5, 2025
    What to do if there is no air circulation in Germfree Laboratory Equipment?
    • S
      Sherry SmithAug 6, 2025
      If there's no air circulation, check the motor blower control. Also, inspect the circuit breaker on the control panel. Finally, using the electrical diagram, check the wiring connections.
  • A
    Amy Walsh MDAug 10, 2025
    What to do if the Germfree BBF-2SSCH Laboratory Equipment air supply is inoperative?
    • D
      Douglas StaffordAug 10, 2025
      If the air supply is inoperative, first, check the main switch to ensure it is on. Make sure the plug is properly inserted into the electrical outlet and that the green light switch is illuminated.
  • E
    Eric ClaytonAug 14, 2025
    What causes exhaust air imbalance in Germfree Laboratory Equipment and how to fix it?
    • M
      Manuel NicholsAug 14, 2025
      Exhaust air imbalance can be caused by several factors. First, check for any obstruction over the exhaust filter and remove it. Second, make sure that the blower motor control is properly adjusted to the correct reading on the Magnehelic gauge. Third, check for and remove any obstruction of the plenum. If these steps do not solve the problem, there may be a leak between the blower and plenum, or the filters may need to be replaced.
  • M
    Michael TorresAug 18, 2025
    Why is the air velocity low in my Germfree BBF-2SSCH?
    • E
      Erika CastilloAug 18, 2025
      Low air velocity could be due to an obstruction in the plenum. Remove any obstruction. Other possible causes include low voltage on house current, loaded filters, improper adjustment of the blower motor, or a leak between the blower and plenum.
  • R
    Rebecca WalkerAug 22, 2025
    What causes high velocity air in Germfree Laboratory Equipment?
    • J
      Jon SanchezAug 23, 2025
      High air velocity can be caused by an exhaust air outlet obstruction. Remove the obstruction. Other possible causes include exhaust filter rupture, supply filter rupture, supply filter gasket leak or improper adjustment of blower motor speed control.
  • M
    Matthew ReedAug 27, 2025
    How to fix excessive blower vibration in Germfree BBF-2SSCH?
    • J
      Jordan BakerAug 27, 2025
      Excessive blower vibration is often caused by foreign matter in the blower squirrel cage. Remove any paper, paper clips, or other debris.
  • M
    Michelle EllisonAug 31, 2025
    What to do if there is non-laminar airflow in Germfree BBF-2SSCH Laboratory Equipment?
    • K
      Kimberly EricksonAug 31, 2025
      Non-laminar airflow can be caused by a large object obstructing airflow. Remove the item.

Summary

Installation

Electrical Services

Control of Airflow

Filters

Certification and Repair

Motor Blower

Performance Testing

Use of the Biosafety Cabinet

Good Technique

Proper procedures for safe and effective cabinet operation.

Biological Safety Cabinet Operation Donts

Troubleshooting