Benchtop Bioflow Chamber (BBF) User Manual
(Class II, Type A/A2)

Questions and Answers

  • A
    Amy RiceAug 1, 2025
    What to do if my Germfree BBF has exhaust air imbalance?
    • S
      smithaliciaAug 1, 2025
      If your Germfree Laboratory Equipment has an exhaust air imbalance, it may be due to several reasons. Check for any obstruction over the exhaust filter, and remove it if present. Ensure the blower motor control is properly adjusted by checking the Magnehelic gauge. Also, inspect the plenum for any obstructions and remove them. Note that if there is a leak between the blower and plenum or the filter is loaded, you should contact a certification company to repair the leak or replace the filters.
  • S
    stevenjenkinsAug 4, 2025
    What to do if my Germfree BBF has low velocity air?
    • V
      Victoria MillerAug 5, 2025
      If the Germfree Laboratory Equipment exhibits low air velocity, there might be several reasons. Check for any obstruction in the plenum and remove it. Note that if there is low voltage on house current, the filter is loaded, there is an improper adjustment of the blower motor, or a leak between the blower and plenum, you should contact a certification company to measure voltage, replace filters, adjust to proper reading on Magnehelic gauge or repair leak.
  • R
    Robert NobleAug 9, 2025
    Why is there no air circulation in my Germfree BBF Laboratory Equipment?
    • S
      Stanley JimenezAug 9, 2025
      If there is no air circulation in your Germfree Laboratory Equipment, it could be due to the blower being off. Check the motor blower control. It could also be due to a blown circuit breaker. Check the circuit breaker on the control panel.
  • J
    Jason HayesAug 13, 2025
    What to do if my Germfree BBF has high velocity air?
    • K
      Keith WilliamsAug 13, 2025
      If your Germfree Laboratory Equipment is experiencing high air velocity, it could be due to several reasons. Remove any obstruction from the exhaust air outlet. Adjust the blower motor speed control to the proper reading on the Magnehelic gauge. Note that if there is an exhaust filter rupture, a supply filter rupture or a supply filter gasket leak, you should contact a certification company to replace filters or repair leak.
  • M
    Misty HoustonAug 18, 2025
    What to do if my Germfree BBF Laboratory Equipment has excessive blower vibration?
    • S
      Susan SmithAug 18, 2025
      If your Germfree Laboratory Equipment has excessive blower vibration, it is likely due to foreign matter in the blower squirrel cage. Remove any foreign objects such as paper or paper clips.
  • M
    Mary CurtisAug 22, 2025
    How to fix non-laminar airflow in Germfree BBF?
    • J
      Jeremy LopezAug 22, 2025
      To fix non-laminar airflow in your Germfree Laboratory Equipment, first check for any large objects obstructing the airflow and remove them. Note that if there is a large leak in supply filter, you should contact a certification company to replace filters.
  • W
    William DaughertyAug 27, 2025
    What to do if the air supply is inoperative in my Germfree BBF Laboratory Equipment?
    • A
      aesparzaAug 27, 2025
      If the air supply is inoperative in your Germfree Laboratory Equipment, check the main switch to ensure it is on. Also, verify that the plug is properly inserted into the electrical outlet and that the green light switch is illuminated.

Summary

Unit Installation and Maneuvering

Transport and Maneuvering Details

Details how to move the unit, its dimensions, and modifications for narrow openings to fit through.

Services and Control Panel Features

Airflow Control and Monitoring

Magnehelic Gauge for Airflow Monitoring

Explains how to use the Magnehelic gauge to monitor airflow, effects of filter loading, and the need for professional certification.

Airflow Diagram and Visualization

HEPA Filters and Maintenance

HEPA Filter Specs and Operational Procedures

Details HEPA filter certification, life, continuous operation, and step-by-step start-up and shutdown procedures.

Motor-Blower System Details

Performance Testing and Verification

Factory and User Performance Tests

Describes factory testing methods (DOP, air velocity) and user-performable tests like smoke testing for air balance.

Operational Guidance and Best Practices

Good Technique and Safety

Operator Hygiene and Work Area Management

Details operator hygiene, proper item placement, and managing the work area for effective containment.

Safety Precautions and Warnings

Critical Operational Donts and Rules

Lists essential prohibitions and safety advice to prevent contamination and ensure operator safety during use.

Troubleshooting Common Issues

Diagnosing Airflow and Circulation Issues

Provides common problems, their causes, and remedies for airflow and circulation issues encountered during operation.

Appendix A: Paraformaldehyde Decontamination

Appendix B: Filter Handling and Replacement

Wiring Diagrams and Schematics

Comprehensive Parts List

BBF Series 220 V Parts and Specs

Germfree BBF Series Specifications

General IconGeneral
TypeBiological Safety Cabinet
ClassII
CertificationNSF/ANSI 49
ConstructionStainless Steel
Class TypeClass II