Germfree - Class II, Type A2 Biosafety Cabinet User Manual - v2.1 - 09SEP2021 - RMH
Class II, Type A2 Biosafety Cabinet
User Manual

Questions and Answers

  • T
    Tony ButlerAug 5, 2025
    What to do if there is no air circulation in Germfree BBF-6SSRX Laboratory Equipment?
    • C
      christina85Aug 6, 2025
      If there's no air circulation, check the motor blower control. Also, inspect the circuit breaker on the control panel. Finally, using the electrical diagram, check the wiring connections.
  • V
    vmilesAug 10, 2025
    What to do if the Germfree BBF-6SSRX air supply is inoperative?
    • M
      Martha ThomasAug 10, 2025
      If the air supply is inoperative, first, check the main switch to ensure it is on. Make sure the plug is properly inserted into the electrical outlet and that the green light switch is illuminated.
  • A
    Ashley MillerAug 14, 2025
    What causes exhaust air imbalance in Germfree BBF-6SSRX and how to fix it?
    • L
      Logan BarnesAug 14, 2025
      Exhaust air imbalance can be caused by several factors. First, check for any obstruction over the exhaust filter and remove it. Second, make sure that the blower motor control is properly adjusted to the correct reading on the Magnehelic gauge. Third, check for and remove any obstruction of the plenum. If these steps do not solve the problem, there may be a leak between the blower and plenum, or the filters may need to be replaced.
  • S
    Sharon BrownAug 18, 2025
    Why is the air velocity low in my Germfree BBF-6SSRX Laboratory Equipment?
    • R
      Richard CooperAug 19, 2025
      Low air velocity could be due to an obstruction in the plenum. Remove any obstruction. Other possible causes include low voltage on house current, loaded filters, improper adjustment of the blower motor, or a leak between the blower and plenum.
  • S
    Stacey GomezAug 22, 2025
    What causes high velocity air in Germfree BBF-6SSRX?
    • R
      rcollinsAug 22, 2025
      High air velocity can be caused by an exhaust air outlet obstruction. Remove the obstruction. Other possible causes include exhaust filter rupture, supply filter rupture, supply filter gasket leak or improper adjustment of blower motor speed control.
  • A
    Ashley SmithAug 27, 2025
    How to fix excessive blower vibration in Germfree BBF-6SSRX Laboratory Equipment?
    • A
      Antonio HenryAug 27, 2025
      Excessive blower vibration is often caused by foreign matter in the blower squirrel cage. Remove any paper, paper clips, or other debris.
  • M
    Mary GreenAug 31, 2025
    What to do if there is non-laminar airflow in Germfree BBF-6SSRX Laboratory Equipment?
    • D
      davismichaelAug 31, 2025
      Non-laminar airflow can be caused by a large object obstructing airflow. Remove the item.

Summary

Installation

Electrical Services

Control of Airflow

Filters

Certification and Repair

Motor Blower

Performance Testing

Use of the Biosafety Cabinet

Biological Safety Cabinet Operation Donts

Troubleshooting

Germfree BBF-6SSRX Specifications

General IconGeneral
TypeBiological Safety Cabinet
AirflowVertical laminar airflow
Width6 feet
ConstructionStainless steel
Filter TypeHEPA
LightingFluorescent
Electrical Requirements120V, 60Hz