4.JN-23 steam sterilizer function and operation instruction
4.1 JN series steam sterilizer contains 1-time(solid/plastic) and 3-times(hollow/prion/cotton) vacuum programs.
With lager color LCD screen for showing pressure, temperature, time, running state and error warning.
Temperature precision: 0.1 ;Pressure precision: 0.001bar.
4.2 Operator interface introduction
Turn on the power at the bottom right,click the power revert button located on the right side of the middle.
Page 16 of 58
SET UP 1
1.Language:
Chinese/Russian/Spanish/Farsi
English/French/Italian
2.Date and time
3.User information
4.Calibration:
Temperature/Pressure/Water quality
monitoring
5.Preference:
6.Service:
Maintenance Infor/Pipeline Cleaning
SystemTest/Alarm Codes/ Auto-Water
Supply
Sleepmode/Delay Start/Power Saving
SET UP 2
1.System information
2.Label Printer
3.USB Record
4.MINI-Printer
5.System Setting
6.Cellphone App
Vacuum Testhot key
134 PROINPROGRAM hot key
Vacuum TestHelix TestB&D Test
Unlock icon.
Only when it is displayed on the screen,
can be open the door.
Main Interface
Sterilization Program:
1.
Solid 134 -04Min-1times
2.Hollow 134 -04Min-3times
3.Prion 134 -20Min-3times
4.Cotton 121 -30Min-3times
5.Plastic 121 -20Min-1times
6.User-definned
Temperature:121 or 134
Sterilization time:04~60Min
Dry time:01~60min
Vacuum times:1times or 3times

Table of Contents