C)when it show Fig.4.3-7 interface warning,means that distilled water was unqualified (σ>80μms).
Prompts you to drain unqualified water in the clean-water tank and add the qualified distilled water again.The warning
Will be eliminated. Water supply requirements:
●conductivity ≤15μs/cm(Room temperature) ●bleach content≤2mg/L
●PH:5~7 ●Hardness:≤0.02mmol/L
Automatic supply water
connection
Drain used-water
connection
Drain clean-water
connection
Highest water level cordon
Fig.4.3-8 Fig.4.3-9
WARN: ,the sterilization program can be started.
Or it will out of operation.
Only if all the warnings were eliminated
D)Automatic water supply and Drain operation (Fig.4.3-8) and adding water manually (Fig.4.3-9)
4.4“SET UP 1” introduction
4.4.1.Language: Chinese/Russian/Spanish/Farsi/English/Franch/Italian
4.4.2.Date and time
4.4.3.User information
4.4.4.Calibration: Temperature/Pressure/Water quality monitoring
4.4.5.Preference:
4.4.6.Service: Maintenance Infor/Pipeline Cleaning/SystemTest/Alarm Codes/ Auto-Water Supply
Sleepmode/Delay Start/Power Saving
Page 18 of 58