2
1. Before installation
The information given in the present manual must by all means be carefully read and
observed. Only then can a perfect functioning of the Water Still be guaranteed.
Vital information within the manual are emphasized in bold letters.
Safety precautions are additionally marked with the following symbols
A free of charge guarantee repair cannot be granted for defects due to improper installation
or handling.
In order to protect the Water Stills during transport, all free openings are covered with tape.
These transport protections have to be removed before installation.
2. Location of the Water Still
The unit is suitable for both bench and wall mounting. It is only suitable for indoor use.
If the unit is to be placed on a bench, place on solid, even and level surfaces only.
If the unit is to be mounted on a wall, check the carrying capacity of the wall in connection
with the total weight of the unit (appliance plus water filling, see technical data).
There are two keyholes for wall mounting at the back of the water still. Place two fastening
screws in the wall in the distance of the two keyholes and nest the still on the keyholes.
Model / Distance of screws
2002 / 40 cm 2004 / 48 cm
2008 / 64 cm 2012 / 64 cm
The unit is not suitable for use in explosion endangered surroundings, eg during
anaesthesia with inflammable gas or steam types!
3. Voltage
Main switch of the water still must be in OFF position.
Mains voltage and voltage stated on the name plate at the left-hand side of the unit
must be identical.
If this applies, unit can be connected.