Caution!
4.6 Tap water inlet (6)
The tap water inlet supplies water to the water still through a solenoid valve. Use a ½"
pressure hose and a blockable water connection (8, hand stop valve) to connect the hose
connection with the water mains.
By all means use hose clips to secure both hose connections!
5. Initiation
5.1 Before initial starting
Before initial starting, the water still has to be hand-filled with water. To fill with water,
remove the outer and the left inner lids and the deflector below the inner lid. Caution, when
working on the deflector, observe the correct position of the wire electrode in the
container. It must neither be bent nor come into contact with the housing. Fill water into the
boiler (left-hand chamber) until the heating elements at the bottom of the container are
below water level. Replace the deflector and the two lids before starting the unit.