EasyManuals Logo

GHM GREISINGER G 1114 Quick Reference Guide

GHM GREISINGER G 1114
22 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #20 background imageLoading...
Page #20 background image
9 | Technical data G 1111 / G 1114
20 / 22 B-H88.0.1X.DK2-1.2
9 Technical data
_
_
_
_
G 1111
Measuring range
0,0 .. 1700,0 hPa (mbar) abs.
0,00 .. 25,00 PSI abs.
0,0 .. 1300 mmHg (Torr) abs.
Overload
Max. ± 3000 hPa abs.
Accuracy
Typ.: ± 2 hPa (at T: 5 .. 30 °C)
Max.: ± 1,25 % FSS
G 1114
Measuring range
0 .. 14000 hPa (mbar) abs.
0,00 .. 199,99 PSI abs.
0 .. 10500 mmHg (Torr) abs.
Overload
Max. ± 20000 hPa abs.
Accuracy
Typ.: ± 0,02 % FSS ± 0,1 % of m.v. (@ 25 °C)
Max.: ± 0,1 % FSS ± 0,5 % of m.v.
Measuring medium
air, clean water or non-corrosive/non-ionising gases
Pressure connection
1 hose connection, interchangeable with G1/8 universal ports
Measuring cycle
FAST: approx. 25 measurements per second
SLO: approx. 2.5 measurements per second
Display
3-line segment LCD, additional symbols, illuminated (white,
duration adjustable)
Standard function
Min/Max/Hold
Activatable special
functions
NVLL: Tare function
AVR: Averaging over 2 s / 5 s / 10 s
Adjustment
Zero point and gradient adjustment

Other manuals for GHM GREISINGER G 1114

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GHM GREISINGER G 1114 and is the answer not in the manual?

GHM GREISINGER G 1114 Specifications

General IconGeneral
BrandGHM
ModelGREISINGER G 1114
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals