EasyManua.ls Logo

Gilson Pipetman Fixed - Troubleshooting

Gilson Pipetman Fixed
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10
8  TROUBLESHOOTING
If there is a leak or the pipette is not accurate,
check for the following:
• Theconnectingnutisloose.
Tighten the connecting nut.
• Thetip-holderiscrackedorscored.
Remove the tip ejector and inspect the tip-holder.
For F2 to F20 models, if the tip-holder is damaged,
the piston may also be damaged. Replace the dam-
aged parts (see Chapter 9). When re-assembling the
pipette, the connecting nut should be hand tightened.
• Chemicaldamagetothepistonandseals.
Replace the piston and seals (see Chapter 9). Wash
the inside of the tip-holder using distilled water.
• Improperreassembly.
Disassemble the pipette and reassemble the pipette
correctly.
If there is liquid in the tip-holder, clean the pipette
as follows:
• Removethetipejector.Unscrewtheconnecting
nut and rinse the tip-holder, piston, seal and O-ring
with distilled water. Dry these parts and reassemble
the pipette.
If necessary, the tip-holder, connecting nut and the
tip ejector can be autoclaved for 20 minutes at 121°C
and at 0.1 MPa pressure. The seal and the O-ring
should not be autoclaved.
If there is an air bubble when the sample is
aspirated:
• Dispensethesampleintoitsoriginalvessel.
• Ensurethatthetip isproperlyimmersedinthe
sample liquid.
• Aspiratethesamplemoreslowly.
If the bubble appears a second time, replace the tip.
The outside of the PIPETMAN Fixed can be cleaned
using a cloth dampened with isopropanol.

Related product manuals