3
The PIPETMAN Fixed can be easily re-adjusted in
the laboratory, with compensation for variations in
uid viscosity and density. A special calibration key
is provided.
The nominal volume of each
model is shown on the small label
on the push-button.
PIPETMAN Fixed is in accordance
with ISO 8655 Standard.
2 SPECIFICATIONS
PIPETMAN Fixed is a high quality pipette that oers
excellent accuracy and precision. The gures given
in the “Gilson Maximum Permissible Errors” table
were obtained using “PIPETMAN DIAMOND Tips”.
These gures are only guaranteed using genuine
PIPETMAN DIAMOND Tips.
The data given in the table conform to the ISO 8655-2
Standard.
Maximum Permissible Errors
Model Volume Gilson ISO 8655
(Reference) (µL) Systematic Random Systematic Random
error (µL) error (µL) error (µL) error (µL)
F2 (F123770) 2 ± 0.08 ≤ 0.03 ± 0.08 ≤ 0.04
F5 (F123771) 5 ± 0.10 ≤ 0.04 ± 0.125 ≤ 0.075
F10 (F123772) 10 ± 0.10 ≤ 0.05 ± 0.12 ≤ 0.08
F20 (F123604) 20 ± 0.20 ≤ 0.06 ± 0.2 ≤ 0.1
F25 (F123775) 25 ± 0.25 ≤ 0.07 ± 0.5 ≤ 0.2
F50 (F123778) 50 ± 0.40 ≤ 0.15 ± 0.5 ≤ 0.2
F100 (F123784) 100 ± 0.80 ≤ 0.25 ± 0.8 ≤ 0.3
F200 (F123605) 200 ± 1.60 ≤ 0.30 ± 1.6 ≤ 0.6
F250 (F123787) 250 ± 3.00 ≤ 0.75 ± 4.0 ≤ 1.5
F300 (F123788) 300 ± 3.50 ≤ 0.75 ± 4.0 ≤ 1.5
F400 (F123789) 400 ± 3.60 ≤ 0.80 ± 4.0 ≤ 1.5
F500 (F123790) 500 ± 4.00 ≤ 1.00 ± 4.0 ≤ 1.5
F1000 (F123606) 1000 ± 8.0 ≤ 1.3 ± 8.0 ≤ 3.0
F5000 (F123607) 5000 ± 30 ≤ 8 ± 40 ≤ 15.0