Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration - electromagnetic immunity
The [KD-558] is intended for use in the electromagnetic environment specied below.
The customer or the user of the [KD-558] should assure that it is used in such
an environment.
IMMUNITY test IEC 60601 Compliance Electromagnetic
test level level environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
± 6 kV
contact
± 8 kV air
3 A/m
Floors should be wood,
concrete or ceramic tile. If oors
are covered with synthetic
material, the relative humidity
should be at least 30%
Power frequency magnetic elds
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment
± 6 kV
contact
± 8 kV air
3 A/m