Thank you for choosing the GlucoMen® areo 2K meter.
We have designed this compact blood glucose and β-ketone meter for
faster and easier testing of your blood glucose and β-ketone we hope it
will be of great help in the management of your diabetes. This manual
explains how to use your new meter. Before testing, carefully read through
this manual and the package inserts that come with it: GlucoMen® areo
Sensor, GlucoMen® areo β-Ketone Sensor, GlucoMen® areo Control, and
GlucoMen® areo Ket Control. Pay particular attention to listed warnings
and cautions. Please keep this manual at hand for future reference. If you
have any questions, please contact A. Menarini Diagnostics for assistance
(references at the end of this manual).
1. Intended use
Your GlucoMen® areo 2K meter, GlucoMen® areo Sensor, and GlucoMen®
areo β-Ketone Sensor are in vitro diagnostic medical devices to
quantitatively measure glucose and β-ketone levels in fresh capillary whole
blood. They are intended for self testing to monitor and control blood
glucose and β-ketone levels by people with diabetes mellitus; they can
also be used in a clinical setting by healthcare professionals. They are not
intended for diagnosis or screening of diabetes and DKA or for neonatal
use. Do not alter your treatment on the basis of test results of this meter
without previously consulting your doctor or healthcare professional.
Blood Glucose and ß-Ketone
Self-Monitoring System
2. System overview
2.1 Meter
FRONT
TEST STRIP PORT
UP/DOWN (
/ )
BUTTONS
ENTER (
) BUTTON
TEST STRIP PORT
Test strip should be inserted here.
ENTER (
) BUTTON
Press and hold
for 2 seconds to access meter memory (§6.1) or longer
(4 seconds) to enter setup mode while not testing (§7).
UP/DOWN (
/ ) BUTTONS
Press
or to scroll through different options and/or values.
BACK
BATTERY HOLDER
NFC ANTENNA
DATA TRANSFER PORT
RELEASE BUTTON
RELEASE BUTTON
Press this button to remove used test strip.
BATTERY HOLDER
Houses two CR2032 3V lithium coin cell batteries.
DATA TRANSFER PORT
Connection port for PC data cable and Bluetooth device from A. Menarini
Diagnostics.
METER SCREEN
CLOCK/ALARM ICON
HYPO/HYPER ICONS
DROP ICON
MAIN DISPLAY AREA
UNIT OF MEASUREMENT
TIME
AVERAGE
DATEMARKERS ICONS
MEMORY ICON
CTL ICON
GLUCOSE ICON
KETONE ICON
WRONG TEMPERATURE
ICON
NFC ICON
LOW BATTERY ICON
MEMORY ICON: indicates that you are using the memory
(§6).
CLOCK/ALARM ICON: clock” appears while setting the time
(§7.1); alarm” appears while setting the
acoustic reminder(s), and will be displayed
if any alarm is on (§7.2).
HYPO/HYPER ICONS: appear while setting the hypoglycaemic
and hyperglycaemic thresholds and if
your test results are below or above these
values (§7.4).
DROP ICON: blinks to indicate that the meter is ready
for blood or control solution testing (§4.2,
4.3).
MAIN DISPLAY AREA: displays test results, stored test values,
result averages and messages.
UNIT OF MEASUREMENT:
shows the unit of measurement of your
blood glucose meter (mg/dL or mmol/L
for glucose, not changeable; mmol/L only
for β-ketone).
USER MANUAL
MASTER
Question and Answer IconNeed help?

Do you have a question about the GlucoMen Areo 2k and is the answer not in the manual?

GlucoMen Areo 2k Specifications

General IconGeneral
Measurement Range20-600 mg/dL
Sample Size0.5 µL
Test Time5 seconds
Blood SampleCapillary whole blood
Operating Conditions10-40°C, 10-90% RH
Battery TypeCR2032
BatteryCR2032

Summary

Intended Use of GlucoMen areo 2K Meter

GlucoMen areo 2K Meter System Overview

Meter Components and Front View

Overview of the GlucoMen® areo 2K meter's front and components like buttons and ports.

Key Buttons and Their Functions

Details the function of the Enter, Up/Down, and Release buttons for meter operation.

Meter Screen Interface Elements

Explains key icons like Drop, Memory, Clock/Alarm, Hypo/Hyper, and display areas.

Data Transfer Port and NFC Antenna

Identifies the data transfer port and NFC antenna for connectivity.

Before Testing Precautions

Meter Usage and Environment Cautions

Important safety and operational precautions for using the meter correctly in various environments.

Test Strip Handling and Storage Cautions

Guidelines for correct usage, handling, and storage of test strips to ensure accuracy.

Measuring Blood Glucose Levels

Blood Sampling Procedures and Safety

Guidance on proper blood sampling techniques and associated safety precautions.

Performing a Blood Glucose Test

Step-by-step instructions for performing a blood glucose test using the meter.

Tips for Accurate Glucose Test Results

Important notes and tips for achieving accurate blood glucose test results.

Glucose Test Result Interpretation Warnings

Warnings and guidance for interpreting "LO" or "HI" glucose test results.

Glucose Control Test Procedure and Warnings

Steps for performing a glucose control test and critical warnings for accuracy.

Marking Glucose Test Results

How to mark test results with pre-meal, post-meal, or exercise markers.

Measuring Blood ẞ-Ketone Levels

Blood Sampling for Ketone Tests

Guidance on blood sampling specifically for ẞ-ketone testing.

Performing a Blood ẞ-Ketone Test

Step-by-step instructions for performing a blood ẞ-ketone test.

Tips for Accurate Ketone Test Results

Important notes for achieving accurate ẞ-ketone test results.

ẞ-Ketone Test Result Interpretation Warnings

Warnings and guidance for interpreting "LO" or "HI" ẞ-ketone test results.

ẞ-Ketone Control Test Procedure and Warnings

Steps for performing a ẞ-ketone control test and critical warnings.

Verifying Acceptable Control Test Results

How to verify if the control test result is within the acceptable range.

Managing Test Results

Importance of Correct Date/Time Settings

Importance of correct time and date settings for accurate result memorization.

Reviewing Glucose Test Results and Averages

How to access and view stored glucose test results and their averages.

Reviewing ẞ-Ketone Test Results

How to access and view stored ẞ-ketone test results.

Data Transmission Methods (NFC)

Methods for downloading test results, including NFC transmission instructions and warnings.

Meter Setup and Configuration

Accessing Meter Setup Menu

Instructions on how to access and navigate the meter's setup menu.

Date and Time Setup Procedure

How to set the date and time on the meter.

Alarm Setup (Glucose and ẞ-Ketone)

How to set up to six alarms for glucose and ẞ-ketone reminders.

Beeper and Hypo/Hyper Alarm Setup

How to configure the beeper and set hypo/hyper glucose alarm thresholds.

ẞ-Ketone Test Warning Setup

How to set a glucose threshold warning for ẞ-ketone testing.

NFC Function Setup

How to enable or disable the NFC function on the meter.

Meter Care and Maintenance

Storage and Cleaning Guidelines

Instructions on proper storage, cleaning, and disinfection of the meter.

Battery Replacement Procedure and Warnings

Step-by-step guide for battery replacement and associated warnings.

Troubleshooting Meter Issues

Understanding HI/LO and Error Messages

Explanation of HI/LO result messages and common meter error codes (Er1-Er5).

Resolving Unexpected Test Results

Steps to take if glucose or ẞ-ketone test results are unusual or inconsistent.

Technical Specifications

General Meter Specifications

General technical details of the GlucoMen® areo 2K meter.

Glucose Measurement Specifications

Specific technical parameters for glucose measurement.

ẞ-Ketone Measurement Specifications

Specific technical parameters for ẞ-ketone measurement.

Product Warranty Information

Symbols and Abbreviations Guide

Related product manuals