MARINE4QUICKSTART,CALIBRATIONANDTROUBLESHOOTINGGUIDE
CalibrationusingABCstation
Equipmentrequired:
1. ABCStationandpoweradapter
2. USBsticksuppliedwithABCstationcompletewithsoftwareas
supplied
3. Testgasandtubingwithconnectors
4. Demandflowregulator
5. Marine4

1. Connectdemandflowregulatorandtubingtogasbottle
2. ConnecttubingintothebackoftheABCstation
3. ConnectAir/ExhaustTubingtoABCStation(ifrequired)
IftheABCStationislocatedinapoorlyventilatedarea,itis
recommendedthattubingisconnectedandroutedtothenearest
windowforthesupplyoffresh
air/exhaustventilation.
4. ConnectpowertoABCstation
5. Allow30secondsfortheunittowarmup,ensurethatthelight
sequenceOrangeOrangeRedisseenduringpowerupafterwhichall
LEDswilllightupgreenandwhencompletedonlythepowerlightwill
belit
5seconds
25seconds
6. Putmemorysticksuppliedinto
PC/laptopandselect“Open
FoldertoViewFiles”(noteactual
displaymayvarywithdifferent
operatingsystems)