GOSSEN Müller & Weigert 13
Select the required values with the and keys. Save values to memory with
the P key. The measuring instrument saves all of the settings to memory and
returns to the normal operating mode.
6 Setting the Limit Values During Operation
Limit values can only be adjusted during operation if they have not been
secured during programming. Press the P and keys simultaneously in order
to enter the limit value adjustment mode. Depending upon the selected
switching function, either Lo-1 and a numeric value, or Hi-1 and a numeric
value blink alternately at the display. The numeric value corresponds to limit
value 1. Select the required value with the and keys. Save the value with
the P key and proceed to the next parameter. Depending upon the selected
switching function, either Lo-2 and a numeric value, or Hi-2 and a numeric
value blink alternately at the display. The numeric value corresponds to limit
value 2.
Select the required value with the and keys. Save value to memory with
the P key. The measuring instrument saves the selected values to memory
and returns to the normal operating mode.
Note: If the instrument has been secured against unauthorized or inadvertent
changes to the limit values, Loc appears at the displays. Adjustment is not
possible in this case.
7 Mean Value Generation
This parameter can only be configured if programming has been enabled. If
programming has not been enabled, Loc appears at the display.
Press the and keys simultaneously.
FILt and a numeral blink alternately at the display.
The numeral indicates the number of measured values which are used for
mean value generation.
Example: 0 = no mean value generation
2 = mean value generation with 2 measured values
Any of the following settings can be used for displayed mean values:
No mean value,
Mean value based upon 2, 4, 8, 16 or 32 measurements.
The instrument automatically selects mean value generation based upon 32
measurements for Pt100 sensors.