EasyManua.ls Logo

GMW DIGEM 96 x 48 B5 - Technical Specifications; Display and Input Characteristics; Error Limits and Accuracy

GMW DIGEM 96 x 48 B5
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14 GOSSEN Müller & Weigert
8Technical Data
Display
Type 7 segment LED
Display Color red, optionally green
Display Range –1999 to 9999
Character Height approx. 13.2 mm
Polarity “–” is displayed automatically
Decimal Point programmable
Overflow Display
_ _ _ _
Input
Module depends on version: see serial plate
Voltage Module
Input Impedance > 1 M for measurements > 2 V
> 70 k for measurements < 2 V
Current Module
Voltage Drop max. 2 V
Temperature Module, Pt100
Sensor Current 2 mA for Pt100
Error Limits
For Basic Instrument
Without Modules ± (0.1% of rdg. + 1 digit)
DC Module ± (0.1% of rdg. + 2 digits)
Temperature Coefficient < 80 ppm/K
SMRR > 35 dB at 50 Hz
CMRR > 120 dB with ref. to MB 200.00 mV at 50 Hz
AC Module (arithmetic)
Intrinsic Error at 45 ... 65 Hz ± (0.2% of rdg.
+ 3 digits)
30 ... 1 kHz ± (0.3% of rdg. + 5 digits)
Temperature Coefficient ± (0.01% + 0.01 mV) / K
TRUE RMS Module
Intrinsic Error at 45 ... 65 Hz ± (0.2% of rdg. + 3 digits)
20 Hz ... 1 kHz ± (0.3% of rdg. + 5 digits)
Crest Factor 6 (additional 0.5% of rdg.)
Temperature Coefficient ± (0.01% of rdg.
+ 0.01 mV) / K
rdg. = measurement value (reading)