GDM-8246 MULTIMETER 
PROGRAMMER MANUAL 
 
⎯  16  ⎯ 
 
:CONFigure:SFRequency  Sets frequency function. 
:CONFigure:RESistance <NR2>  Sets resistance function and range. 
:CONFigure:VOLTage:AC <NR2>  Sets AC voltage function and range.
:CONFigure:VOLTage:DC <NR2>  Sets DC voltage function and range.
:CONFigure:VOLTage:ACDC <NR2>  Sets AC+DC voltage function and 
range. 
:CONFigure:VOLTage:DCAC <NR2>  Sets Ripple voltage function and 
range. 
:READ?  Returns the value displayed on the 
primary and secondary display 
:VALue?  Returns the value displayed on the 
primary display. 
:SVALue?  Returns the value displayed on the 
secondary display. 
  GDM-8246 MULTIMETER 
PROGRAMMER MANUAL 
 
⎯  17  ⎯ 
z 
Calculation Commands 
:CALCulation: LIMit:STATe?  Returns Compare mode on or off. 
:CALCulation: LIMit:STATe <Boolean>
 
Sets Compare mode on or off. 
:CALCulation: LIMit:LOWer? 
Returns the value of the lower 
limit. 
:CALCulation: LIMit:LOWer <NR2> 
Sets the value of the lower limit. 
:CALCulation: LIMit:UPPer? 
Returns the value of the upper 
limit. 
:CALCulation: LIMit:UPPer <NR2> 
Sets the value of the upper limit. 
:CALCulation: LIMit:FAIL? 
Returns the limit result. 
:CALCulation:MAXimum?
 
Returns the MAX mode on or 
off. 
:CALCulation:MAXimum <Boolean>
 
Sets the MAX mode on or off. 
:CALCulation:MINimum?
 
Returns the MIN mode on or off. 
:CALCulation:MINimum <Boolean>
 
Sets the MIN mode on or off. 
:CALCulation: RELation: STATe <NR1>
Sets REL mode on or off. 
:CALCulation:RELation: STATe?
 
Returns REL mode on or off. 
:CALCulation:RELation:DATa?
 
Returns the reference value of 
the REL mode. 
:CALCulation:RELation:DATa <NR2>
 
Sets the reference value of the 
REL mode.