MAIN 
OUTPUT 
Connect an  oscilloscope to output. 
Observe  a 20Vp-p  20KHz  (approx.)  triangle wave. 
FUNCTION 
SWITCH 
Select and observe a 
20V 
p-p  square wave  and  sine wave. 
AMPLITUDE 
CONTROL 
Rotate the AMPLITUDE  vernier from maximum  to minimum  and 
observe greater than 
23dB 
of  attenuation. 
ATTENUATION 
Connect  the oscilloscope to the OUTPUT  and  push  the 
ATT 
pushbutton  switch that the signal 
is 
attenuated 
by 
a  factor of 
2 
0dB. 
DC 
OFFSET 
Reconnect  the oscilloscope to the 
OUTPUT 
and  select the triangle 
waveform.  Rotate the 
DC 
OFFSET  control potentiometer  (pull position) 
and observe the peaks  of  the triangle waveform 
will 
"clip" when  the 
DC 
OFFSET  plus the peak  amplitude exceeds 
Z10V. 
Reduce  the output amplitude  and  observe the amount  of 
DC 
OFFSET 
may  be  increased by  the same  amount  the peak  amplitude has been 
decreased. 
Return the potentiometer to "PUSH"  position and  the 
AMPLITUDE 
to 
maximum.