OPERATION
55
Setting PASS HOLD
Background
The PASS HOLD settings only apply to the
selected test in an AUTO test. When the PASS
HOLD setting is set to ON, a PASS judgment is
held until the START button is pressed.
Note
The PASS HOLD setting only applies to AUTO
tests. This setting is ignored when running a
single MANU test.
Steps
1. Press the UTILITY key on the front
panel when the tester is in EDIT
status. The display will go from the
normal EDIT status to the MANU
Utility menu for the current test.
UTILITY
L I T Y IM A UN U T 2M A N U = * * * - 0 0
A R C M D E : OO FF
G R O U N D MO D E : NO
P A S S H L D :O O FF
F A I L M D E :O
PS T O
M A X H O L D : O F F
Note
The MANU UTILITY settings only apply to the
selected MANU test.
2. Use the UP/DOWN arrow keys to
move to the PASS HOLD setting.
3. Use the scroll wheel to set PASS
HOLD.
PASS HOLD
OFF, ON
4. Press the EDIT/SAVE key to save
and exit the MANU Utility menu.
EDIT/SAVE
w ww . . co m
information@itm.com1.800.561.8187

Other manuals for GW Instek GPT-9804

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-9804 and is the answer not in the manual?

GW Instek GPT-9804 Specifications

General IconGeneral
BrandGW Instek
ModelGPT-9804
CategoryTest Equipment
LanguageEnglish

Related product manuals