OPERATION
103
The Interlock function is a safety feature. The
interlock function prevents a test from running,
unless the interlock pins on the signal I/O port
connector are shorted. The included interlock
key can be used for this purpose. See page 110
for details.
Steps
1. Ensure the tester is in VIEW status.
Save the current test if necessary.
Page 63
I R G B
E
A CW D CW
F R E Q = 0 H z 6
0
100
k V
EIV
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
W
MM A UN N A_0M A N U = * * * - 0 0
m A
VIEW status
2. Press the UTILITY key.
UTILITY
3. Press the CTRL soft-key to bring
up the Control Common Utility
menu.
T R LC
K e y L o k : Oc
F F
S t a r t t r lC
D o u b l e A c t
i no
L C D B U Z Z N T EI R C T R L
:
: F R O N T P A N
MOC MO
U N T I L I T Y
E L
FO F
I N T E R L C K :O
O FF
4. Use the UP/DOWN arrow keys to
choose a menu item: Start Ctrl,
Double Action, Key Lock or
INTERLOCK.
5. Use the scroll wheel to select
setting for the chosen menu item.
w ww . . co m
information@itm.com1.800.561.8187

Other manuals for GW Instek GPT-9903

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-9903 and is the answer not in the manual?

GW Instek GPT-9903 Specifications

General IconGeneral
BrandGW Instek
ModelGPT-9903
CategoryTest Equipment
LanguageEnglish

Related product manuals