EasyManuals Logo

GW Instek GPT-9903 User Manual

GW Instek GPT-9903
182 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #107 background imageLoading...
Page #107 background image
EXTERNAL CONTROL
107
Remote Controller Operation
Description
The GPT-9000/9000A accepts external remote
controllers with a START and STOP button. To
use the REMOTE terminal, the
GPT-9000/9000A must first be configured to
accept a remote controller.
Operating a remote controller is the same as
operating the START and STOP buttons on the
front panel.
Steps
1. Insert the lead of remote controller
into the REMOTE terminal.
REMOTE
2. Configure the Start Ctrl option to
REMOTE CONNECT in the
Common Utility menu.
Page 102
3. The tester will now only be able to
start a test using a remote
controller.
NOTE
Even if the GPT-9000/9000A is configured to use
the REMOTE CONNECT option, the STOP button
on the front panel can still be used to stop a test.
4. To return the operation control to
the front panel, configure the Start
Ctrl option to FRONT PANEL.
Page 102
w ww . . co m
information@itm.com1.800.561.8187

Other manuals for GW Instek GPT-9903

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-9903 and is the answer not in the manual?

GW Instek GPT-9903 Specifications

General IconGeneral
BrandGW Instek
ModelGPT-9903
CategoryTest Equipment
LanguageEnglish

Related product manuals