GPT-9000/9000A Series User Manual
42
Setting the Test Voltage or Test Current
Background
Steps
1. Press the UP / DOWN arrow keys
to bring the cursor to the voltage
setting.
I R G B A CW D CW I / LH O I M ET R
E
F R E Q = 0 H z 6
0
100
k V
IDE
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
T
2M A N U = * * * - 0 0
cursor
m A
2. Use the scroll wheel to set the
voltage level.
ACW
DCW
IR
GB
0.050kV ~ 5kV
1
0.050kV ~ 6kV
2
0.05kV ~ 1kV (50V steps)
3
3.00A ~ 33.00A (GPT-98XX)
3.00A~ 33.00A (GPT-99XX)
1
At least 0.5 seconds is needed to reach a set
voltage of 50V/10mA.
2
At least 0.5 seconds is needed to reach a set
voltage of 50V/2mA.
3
GPT-99XX/99XXA includes a 125V test point.