SFG-2000 Series User Manual 
28 
Set the Duty Cycle (Square Waveform) 
The duty cycle setting is not available in sine/triangle waveform. 
DUTY
7
   
Press the Shift key, then 7 to 
enter duty cycle editing mode. 
The duty sign appears on the 
display. 
Enter duty 
cycle 
 
The default value is 50%. 
The settable range is 20% ~ 
80%, 2Hz ~ 1MHz. 
 
54
Hz/%
 
Use numerical keys to enter 
value. 
Example: 45% 
 
MHz
 
Delete a number (backspace) 
Edit duty 
cycle 
 
The editing knob changes the 
value, and the cursor keys 
moves the active digit (same as 
entering frequency). 
 
When inactive for 5 seconds, the display automatically 
goes back to previous mode (frequency view). 
 
Setting duty cycle (press 
DUTY
7
) is not available 
when sine or triangle waveform is activated. The 
following message appears. 
 
For full error message list, see page
52. 