GETTING STARTED
19
Rear Panel
AC Voltage
Selector
AC Power
Input
External
Modulator
Input
External
Modulator Input
Accepts the modulation signal from external device.
BNC male connector, 10Vp-p maximum. SFG
automatically switches the modulation signal from
internal to external.
For modulation details, see page
37 (AM) or page38(FM).
AC Power Input
Accepts the AC power cord. 115 or 230V, 50/60Hz.
AC Voltage
Selector
Selects 115V or 230V for power source.
AC 100/110/120V select 115V.
AC 220/230/240V select 230V.
Improper selection might lead to internal fuse
blowout.