Questions and Answers

  • T
    Taylor DavisSep 10, 2025
    What does W20 Measurement process incomplete mean for HACH LANGE ORBISPHERE 6110?
    • M
      Mark HerringSep 10, 2025
      If your HACH LANGE Measuring Instruments display warning W20, it means the measurement process is incomplete because of insufficient initial package pressure. This can happen if the pressure in the package is too low to continue with a measurement, so you should restart the measurement with another package.
  • L
    lsmithAug 22, 2025
    What to do if HACH LANGE ORBISPHERE 6110 Measuring Instruments show E1 Emergency STOP error?
    • R
      Robert JamesAug 23, 2025
      If you see error E1 on your HACH LANGE Measuring Instruments, it means the emergency STOP button has been pushed. To fix this, release the emergency STOP button by turning it clockwise. If necessary, initialize the motor.
  • R
    rlambAug 25, 2025
    Why HACH LANGE Measuring Instruments show E100 Ultrasound error?
    • Y
      yrodriguezAug 25, 2025
      If your HACH LANGE Measuring Instruments display error E100, it indicates an ultrasound error. This can happen because the ultrasound power supply is OFF, so you should check the ultrasound power supply. It can also happen if the emergency STOP button is not released, so you should release the emergency STOP button by turning it clockwise. Finally, too much water in the ultrasound generator can cause this error, so remove any excess water.
  • D
    Don PerryAug 26, 2025
    What causes E2 Motor error on HACH LANGE ORBISPHERE 6110?
    • K
      Krista HoAug 26, 2025
      If your HACH LANGE Measuring Instruments display error E2, it indicates a motor error. This can happen if the measuring head has missed an obstacle, so you should initialize the motor. This can also happen if the emergency STOP button pushed if the motor is moving.
  • K
    Kimberly PowellAug 27, 2025
    What to do if HACH LANGE ORBISPHERE 6110 show E400 error?
    • C
      Christina McdonaldAug 28, 2025
      If you are seeing error E400 on your HACH LANGE Measuring Instruments, it means there was a problem initializing the motor displacement, and the measurement was aborted. To solve this, turn the instrument off and restart it.
  • W
    wsteeleAug 29, 2025
    Why HACH LANGE ORBISPHERE 6110 show E4 error?
    • A
      Amy DavisAug 29, 2025
      If your HACH LANGE Measuring Instruments display error E4, it means the front door has been opened, stopping the measurement process. This can happen if the front door has been opened before the end of the measurement process, so you should close the front door. It can also happen if the front door is not closed properly, so you should control the front door mechanism.
  • B
    Brett EricksonSep 18, 2025
    What does W80 Measurement process incomplete mean for HACH LANGE Measuring Instruments?
    • R
      Renee FerrellSep 18, 2025
      If your HACH LANGE Measuring Instruments display warning W80, it means the measurement process is incomplete because of insufficient package pressure during the ultrasound stage. This can happen if the measured pressure is too low to compute the second phase of the measurement, so you should check foam intrusion in the gas path.
  • D
    delacruzangelSep 12, 2025
    What to do if HACH LANGE ORBISPHERE 6110 show W40 Spontaneous degassing warning?
    • S
      Sandra MillerSep 12, 2025
      If your HACH LANGE Measuring Instruments display warning W40, it means there was spontaneous degassing... This can happen if a spontaneous degassing is detected at the end of the first expansion, so you should restart the measurement with another package.
  • T
    Tamara MendezSep 8, 2025
    Why HACH LANGE ORBISPHERE 6110 show E40 Check gas supply error?
    • V
      vpotterSep 8, 2025
      If your HACH LANGE Measuring Instruments display error E40, it means to check the gas supply, so you should check the purge gas pressure because before filling the reference volume, the pressure is lower than 1.5 bar.
  • C
    Clinton VasquezSep 5, 2025
    What causes E20 Not enough pressure error on HACH LANGE Measuring Instruments?
    • W
      Wesley HallSep 5, 2025
      If your HACH LANGE Measuring Instruments display error E20, it means there is not enough pressure in the system and the measurement process has been canceled. This can happen because of a problem of flow regulation during the measurement, so you should check the purge gas pressure and change the Goretex filter. It can also happen if the gas path is partially or totally blocked, so you should control the antifoam injection.

Summary

General Information

Safety Information and Precautions

Covers hazard symbols, safety precautions, specific warnings, and service policies.

Instrument Specifications

Technical Specifications

Detailed specifications for analysis range, units, operating limits, enclosure, and compliance.

Instrument Installation

Installation Procedures

Guides through pre-installation, physical installation, and post-installation steps.

User Interface and Startup

Instrument Startup Process

Details the automatic component check performed at startup and user login procedures.

Instrument Configuration

Units, Resolutions, and Parameters

Procedures for defining units, resolutions, package parameters, and formulas.

Instrument Calibration

Sensor Calibration Procedures

Step-by-step guides for calibrating barometric, pressure, temperature, CO2, flow, and O2 sensors.

Performing Measurements

Instrument Maintenance

Troubleshooting Guide

Error and Warning Messages

Lists error codes, warning messages, causes, solutions, and startup LED meanings.

Measurement Problems

Common measurement issues, their explanations, and troubleshooting steps.

HACH LANGE ORBISPHERE 6110 Specifications

General IconGeneral
ParameterDissolved Oxygen, Ozone, Hydrogen Peroxide
Accuracy (Dissolved Oxygen)± 0.1 mg/L
IP RatingIP65
Range (Dissolved Oxygen)0 to 50 mg/L
Range (Ozone)0-20 mg/L
Range (Hydrogen Peroxide)0 - 2 ppm; 0 - 20 ppm
Response time (t90)< 60 seconds
Operating temperature0 to 50 °C
Operating pressure0 to 6 bar
Wetted materialsStainless steel, silicone
Sensor typeOptical
CommunicationRS-485
DisplayNo
Outputs4-20 mA, Digital
TypePortable sensor