LISTING DEVICE SPECIFICATIONS
The approximate weight and dimensions of the PCS2 device are as follows:
The following environmental conditions should be respected pertaining to oper-
ation and storage of the PCS2 device:
The electrical specifications for operating the PCS2 device are as follows:
Note: Haemonetics will regulate the proper voltage setting upon installation.
The power source used must be properly grounded.
Characteristics
Values
Cabinet
cover open
Cabinet
cover closed
Height 63 cm 44 cm
Width 55 cm
Depth 55 cm 32 cm
Depth with communication box 55 cm 37 cm
Weight 26.4 kg
Weight with communication box 27.4 kg
Conditions Values
Ambient operating temperature +18° C to +27° C
Tested storage temperature 0° C to + 40° C
Storage humidity level Maximum relative humidity rate of 90%,
non-condensing
Characteristics Values
(relative to input voltage)
Input voltage 230 VAC ± 10% 110 VAC ± 10%
Operating current ~1.9 A ~ 2.6 A
Fuse rating F2.5 A @ 250 V F5.0 A @ 250 V
Operating frequency range 50 - 60 Hz 50 - 60 Hz
Maximum leakage current 500 µA 100 µA