Parameter
code
Parameter meaning
Parameter
code
Parameter meaning
L0 Control temperature setting L3 Current voltage inquiry
L1 High temperature alarm value setting L4 Current environment temperature inquiry
L2 Low temperature alarm value setting L5 Non-use setting
U disk data reading method
U disk interface is on the right of refrigerator-freezer top display board. After inserting U disk to
interface, cabinet will automatically identify and import temperature data. Meanwhile press setting
and temperature area switch key on display board for 5s, check whether data is imported to U disk.
If not, display screen will show "USB" "……"; or it will show "USB""OK". Unplug U disk after
importing, and open the data in computer by text or excel. This product can save the cabinet
temperature data in recent 10 years, if it is more than 10 years, the previous data will be replaced
automatically.
Temperature setting
14
15
1. Unlock
When the lock indicator lights, all keys are locked except digits switch key, it should be unlocked
before setting parameters, the unlock steps are as follows:
Long press digits switch key for more than 5s, the display area shows F0, 0 digit flashing;
Press temperature adjusting key, display area shows F1;
Press setting key, lock indicator goes out, and the keyboard is unlocked.
2. User parameter setting
User can change the setting temperature, alarm value, start delay and other parameters of
refrigerating chamber or freezing chamber through control panel. These settings are independent for
refrigerating chamber and freezing chamber. The operations as refrigerating chamber setting will be
shown on the display screen of refrigerating chamber on left side of display panel, and the parameter
adjustment of freezing chamber will be shown on the refrigerating chamber on right side of display
panel. The activated setting area will make blinking display, to activate another setting area, you
can press temperature switch key.
① Parameter inquiry and setting
Press setting key under unlock status, the display screen activated by default will show L0, 0 digit
flashing, then the temperature area can be changed by temperature area switch;
Press temperature adjustment key to adjust L0 to L1, L2, L3, L4, L5;
Adjust to designated parameter, and press setting value, the display screen will show the setting
value of that area, and the value at units' digit will flash, the corresponding list of parameter code is
as follows:
Press digit switch key to change the flashing digit, press temperature adjustment key to change
current digit value;
Press setting key when adjusting to required temperature, change and save parameters.
The setting temperature by default of cold storage area is 5°C, the frozen area is -40°C;
For non-use setting, the default value is 1, meaning the normal operation of current activated area,
0 means the current activated area is disabled;
Voltage and environment temperature can only be inquired, but can't be set .
High-low temperature alarm value can be set within the following range:
The highest setting temperature+5°C≥high temperature alarm value≥setting temperature+5°C
The lowest setting temperature-5°C≤low temperature alarm value≤setting temperature-5°C
The “├” in front of figure represents “+”.
② Start delay setting
In unlock status, press setting key and temperature adjustment key for 5s, the display screen
activated by default will show the delay time of current area compressor, then the temperature
area can be changed by temperature area switch;
Press temperature adjustment key to change start delay time;
Press setting key when adjusting to required value, change and save parameters.
③ Freezing chamber defrost
In unlock status, press freezing defrost for more than 5s, the refrigeration system of freezing
chamber will stop running, the display screen of cold storage area will show dF, it can't make auto
reply in this status, but long press freezing defrost key under unlock status again for more than 5s to
exit defrost mode, dF will disappear from display screen of cold storage area after successful exit.
In all parameter adjustment, if there is no operation within 10s, system will save
current setting and exit parameter setting mode;
For parameter adjustment, frozen area setting is activated by default for first
use, if it needs to switch to cold storage area setting, press temperature switch
key. If it is not for first use, the final set temperature area is activated by default;
Application method