11
Under range: A blinking “0.00” indicates that the sample
absorbs less light than the zero reference. Check the procedure
and make sure you use the same cuvet for reference (zero) and
measurement.
Over Range: A flashing value of the maximum concentration
indicates an over range condition. The concentration of the
sample is beyond the programmed range: dilute the sample and
re-run the test.
c) during calibration procedure
Standard Low: The standard reading is less than expected.
Standard High: The standard reading is higher than expected.
No Light:The instrument cannot adjust the light level. Please
check that the samples does not contain any debris.
Inverted cuvets: The sample and the zero cuvet are inverted.
Zero: A zero reading was not taken. Follow the instructions of
the measurement procedure for zeroing the meter.