18
• After a few seconds, the display will show
“-0.0-”. The meter is now zeroed and
ready for measurement.
Remove the cuvet from the instrument.
Remove the cap.
Add the content of the one packet of
HI 95771A-0 Chlorine Ultra HR reagent A.
Add the content of the one packet of
HI 95771B-0 Chlorine Ultra HR reagent
B. Replace the cap and shake gently for
20 seconds.
Replace the cuvet into the cuvet holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press and hold READ/
/TIMER for
three seconds. The display will show the
countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 1 minute and just
press READ/ /TIMER.
In both cases, the lamp, cuvet and detector
icons will appear on the display,
depending on the measurement phase.
At the end of measurement, the instrument
directly displays concentration in mg/L of
free chlorine and the range number on
the LCD.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI 96771C and is the answer not in the manual?

Hanna Instruments HI 96771C Specifications

General IconGeneral
BrandHanna Instruments
ModelHI 96771C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals