The BPM18 Wrist Type Blood Pressure Monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of the BPM18 Wrist Type Blood Pressure
Monitor should assure that it is used in such an environment.
Guidance and manufacturer's declaration-electromagnetic immunity
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic
environment-guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
± 6 kV
contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative
humidity should be at least 30 %.
± 6 kV
contact
± 8 kV air
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
3 A/m
21
EMC Tables