Questions and Answers

  • S
    Stephanie MendozaAug 8, 2025
    What to do if aerosol is visible but not coming out of the Hikoneb Respiratory Product tubing?
    • A
      Allen DennisAug 8, 2025
      If you observe aerosol production but it's not exiting the tubing of your Hikoneb Respiratory Product, the issue could stem from several causes. First, ensure the patient tube isn't blocked due to incorrect positioning; reposition it to point upwards and straight. Second, the filter may be occluded, in which case, replace it. Third, verify that the Flow setting is activated in the menu. Finally, a faulty fan could be the culprit, requiring a fan replacement.
  • J
    Joe FranklinAug 13, 2025
    Why is the red error LED on my Hikoneb Respiratory Product?
    • C
      Cody LesterAug 13, 2025
      If the red error LED is illuminated on your Hikoneb Respiratory Product, it typically indicates a low water level or that the nebulizing chamber isn't correctly locked. To resolve this, first, add more water to the device. Second, ensure the chamber is properly fitted and correctly locked onto the device.
  • R
    Robert PierceAug 16, 2025
    What to do if On-Off button is on but there is no aerosol production in Hikoneb Respiratory Product?
    • L
      lhardyAug 16, 2025
      If the On-Off button is illuminated on your Hikoneb Respiratory Product, but there is no aerosol production, the cause may be a faulty transducer or an open circuit (O/C) fuse. To address this, first, replace the transducer. Second, replace the fuses.
  • T
    Travis SmithAug 21, 2025
    How to troubleshoot a Hikoneb Respiratory Product that is not operational?
    • J
      Jeffrey HansonAug 21, 2025
      If your Hikoneb Respiratory Product isn't working and the On-Off button light is off, it's likely due to a power supply issue. Check the plug, IEC connections, and fuses to ensure they are properly connected and functioning.
  • M
    Michael BrownAug 26, 2025
    What to do if aerosol production is visible but not coming out of tubing in Hikoneb 908 DC Medical Equipment?
    • M
      Mr. Eric HullAug 26, 2025
      If you observe aerosol production but it's not exiting the tubing, the issue could stem from several causes: * Incorrect positioning of the patient tube. * A blocked filter. * Deactivated Flow setting. * A faulty fan. To resolve this, try the following: * Reposition the aerosol tube so it points upwards and straight. * Replace the filter. * Activate the Flow setting. * Replace the fan.
  • K
    Kristin HarrisAug 29, 2025
    What to do if the red error LED is on in Hikoneb 908 DC?
    • D
      David LaraAug 29, 2025
      If the red error LED is illuminated, it indicates that either the water level is low or the nebulizing chamber isn't correctly locked. To fix this: 1. Add more water. 2. Re-fit the chamber, ensuring it is correctly secured.
  • S
    Shane MartinezSep 2, 2025
    What to do if On-Off button is on but there is no aerosol production in Hikoneb 908 DC?
    • S
      Stephen BairdSep 3, 2025
      If the On-Off button is illuminated but there's no aerosol being produced, the problem might be due to: 1. A faulty transducer. 2. An open circuit (O/C) fuse. To address this: 1. Replace the transducer. 2. Replace the fuses.

Hikoneb 908 DC Specifications