- 34 -
Overfilled pipette - Safety valve is activated - Remove safety valve, clean and re-assemble.
immediately and prevents liquid If required, replace the safety valve. During
from getting into the the unit assembly take note of point 8.3.
Suction too slow or too - Selected stage is too low - Improve the selection to a greater stage
weak e.g. 3 -> 5.
- No tight, secure connection - Screw the pipetholder more tightly onto the
of the pipetholder handle.
- Filter is full - Exchange the filter.
- Silicone adaptor is damaged - Exchange the silicone adaptor.