EasyManuals Logo

Honeywell DR4300 Replacement Instructions

Honeywell DR4300
18 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
3/05 Replacement Kit Instruction 43-DR-33-87 15
Procedure for Pen Alignment (Recorder with Display)
Table 8 gives the procedure for aligning pens at zero and span (full scale) in a recorder with a
display.
Table 8 Procedure for Pen Alignment (Recorder with Display)
Step Press Action/Result
1
SET
UP
until you can see:
ALIN
Upper Displa
y
PEN
Lower Displa
y
2
FUNC
until you can see:
(value)
Upper Displa
y
(approximately 60 on the upper display)
PEN0
Lower Dis
p
la
y
Wait until pen stops moving downscale.
If the pen tip is greatly out of position, follow Steps 3 through 6.
If the pen position requires only minor adjustment, go to Step 7.
3
Turn off the power to the recorder.
4
To place the pen tip in approximately the correct position,
loosen the screw holding the pen 1 (purple) arm to the servo
shaft.
5
Align the tip of pen 1 with the zero line on the chart and tighten
the locking screw.
6
Turn on the power and repeat Step 2.
7
or
to set tip of pen 1 (purple) on zero scale line of chart ±0.25% (or
quarter graduation).
8
FUNC
and you will see:
(value)
Upper Displa
y
(approximately 1200 more than the zero value in
Step 2)
PEN 100
Lower Display
Wait until pen stops moving upscale.
9
or
to set tip of pen 1 (purple) on full scale line of chart ±0.25% (or
quarter graduation).
10
DISP
to return to normal operation for pen 1. You will see:
(value)
Upper Display
INP 1
Lower Display

Other manuals for Honeywell DR4300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell DR4300 and is the answer not in the manual?

Honeywell DR4300 Specifications

General IconGeneral
BrandHoneywell
ModelDR4300
CategoryVoice Recorder
LanguageEnglish

Related product manuals