Model
1740A
Operation
SECTION
Ill
OPERATION
3-1.
INTRODUCTION.
3-2.
This
section
provides
general
operating
instruc-
tions
for
Model
1740A.
Front-
and
rear-panel
controls
and
connectors
are
identified
and
described
in
figure
3-1.
An
initial
turn-on
procedure,
operators
calibration,
trigger
selection
table,
and
procedures
for
obtaining
basic
displays
are
also
included.
The
index
numbers
after
control
and
connector
names
in
the
text
are
keyed
to
figure
3-1.
3-3.
TURN-ON
PROCEDURE.
Before
turning
on
the
oscilloscope,
read
the
safety
summary
located
at
the
front
of
this
manual.
3-4.
To
turn
on
the
Model
1740A,
perform
the
fol-
lowing
steps:
a.
Turn
all
control
knobs
to
the
12
o'clock
posi-
tion
except
verniers
@
and
SWEEP
VERNIER
should
be
in
CAL
position;
TRIGGER
HOLDOFF
on
MIN.
MAIN
TIME/DIV
@
fully
clockwise.
b.
Verify
pushbuttons
out
except
A@,
AQ,
and
MAIN
@
c.
Press
LINE
switch
@;
LINE
indicator
@
should
light.
After
CRT
warm
up,
a
free-running
trace
should
be
observed
near
center
screen.
d.
Increase
(or
decrease)
BEAM
INTENSITY
@
to
comfortable
viewing
level,
and
adjust
FOCUS
5
|
as
necessary
for
sharpest
trace.
3-5.
OPERATOR
CHECKS.
3-6.
A
few
checks
and
adjustments
may
be
required
to
verify
that
the
Model
1740A
is
operating
properly.
If
the
oscilloscope
is
moved
from
one
electromagnetic
environment
to
another,
the
trace
alignment
control
may
need
adjustment
to
align
the
horizontal
trace
with
the
graticule.
Astigmatism
and
focus
controls
may
need
adjustment
to
obtain
the
sharpest
display.
Probe
compensation
may
be
required,
since
total
input
resistance
and
capacitance
can
vary
slightly
from
one
oscilloscope
to
another.
3-7.
Adjust
trace
alignment
as
follows:
a.
Obtain
a
display
as
described
in
the
turn-on
procedure.
b.
With
vertical
POSN
control
@,
align
trace
with
center
graticule
line.
c.
With
a
screwdriver,
adjust
TRACE
ALIGN
@
(on
rear
panel)
for
best
trace
alignment
with
graticule
line.
3-8.
Adjust
astigmatism
and
focus
as
follows:
a.
Select
A
VS
B
@
and
lower
BEAM
INTEN-
SITY
@
to
a
low
level.
b.
Position
spot
near
center
of
CRT
with
POSN
@
and
POSITION
@
controls.
c.
Adjust
FOCUS
5
]
and
ASTIGMATISM
©
(on
rear
panel)
for
the
smallest
round
spot.
3-9.
Perform
probe
compensation
adjustment
as
follows:
a.
Connect
probe
to
be
compensated
to
appro-
priate
vertical
INPUT
connector
21)
and
the
CAL
1
V
output
@.
b.
Set
VOLTS/DIV
@
to
0.1,
MAIN
TIME/DIV
©
to
0.2
mSEC,
and
input
coupling
@
to
DC.
c.
Adjust
main
TRIGGER
LEVEL
@
fora
stable
display
of
the
calibrator
square-wave
voltage.
Display
should
have
flat
tops.
Any
distortion
in
presentation
is
caused
by
incorrect
probe
compensation.
d.
If
overshoot
or
undershoot
is
present,
turn
screwdriver
adjustment
in
probe
for
a
flat-top
pre-
sentation
(see
figure
3-2).
3-10.
Perform
vertical
accuracy
check
as
follows:
a.
Set
controls
to
positions
indicated
in
turn-on
procedure.
b.
Apply
CAL1V
®
signal
to
channel
AINPUT
connector
using
a
BNC
to
banana
plug
adapter
and
a
test
lead
with
alligator
clips.
c.
Adjust
channel
A
VOLTS/DIV
@
to
0.2
V/div
and
MAIN
TIME/DIV
@
to
0.2
mSEC/div.
Square-
3-1