EasyManua.ls Logo

HP 333A - Section V: Maintenance; Fundamental Rejection Check

HP 333A
53 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TM 11-6625-1576-15
Section V
Paragraphs 5-10 to 5-11, Figure 5-2 and Table 5-2
setting.
Each decrease of the RANGE
switch represents 10 db fundamental re-
jection. Therefore, the total fundamental
rejection is the sum of the Wave Analyzer
reading and the 333A/334A indication.
k. The Distortion Analyzer METER RANGE setting
plus the Wave Analyzer RANGE setting plus the two
meter indications shall total more than -80 db.
5-10. SECOND HARMONIC ACCURACY CHECK.
a. Connect Test Oscillator (-hp-Model 651A) 600
output to Distortion Analyzer *
b. Set Distortion Analyzer controls as follows:
FUNCTION Selector . . . . . . . SET LEVEL
FREQUENCY RANGE . . . . . . . . . . . X1
Frequency Dial . . . . . . . . . . . . . . .20
METER RANGE Selector . . . . . . . . . 0 DB
c. Set Test Oscillator controls (-hp- Model 651A)
as follows:
FREQUENCY RANGE . . . . . . . . . . . X10
FREQUENCY Dial. . . . . . . . . . . . . 2
OUTPUT ATTENUATOR . . . . . 1.0 VOLTS
AMPLITUDE . . . . . . . . . . . . . . l Volt
d. Adjust Distortion Analyzer SENSITIVITY con-
trols for a meter reading of O db.
e. Switch Distortion Analyzer FUNCTION selector
to DISTORTION. Adjust Frequency dial and BALANCE
controls for a null indication on meter.
f. Switch Distortion Analyzer FUNCTION selector
to SET LEVEL.
g. Adjust Test Oscillator frequency to 40 cps. Ad-
just AMPLITUDE control for a O db indication on the
Distortion Analyzer meter.
h. Switch Distortion Analyzer FUNCTION selector
to DISTORTION. The meter reading shall not change
more than ±0. 6 db.
Model 333A/334A
j. Repeat steps a through h at the frequency settings
indicated in Table 5-2.
The meter readings shall
change within the limits specified.
Table 5-2. Second Harmonic Accuracy Check
5-11.
a.
b.
c.
DISTORTION INTRODUCED BY INSTRUMENT
CHECK AND AUTOMATIC CONTROL LOOP
OPERATION.
Connect 333A/334A as shown in Figure 5-2.
Set Distortion Analyzer controls as follows:
FUNCTION Selector . . . . . . . SET LEVEL
SENSITIVITY Selector
. Position 1 step CCW
from Full CW position
SENSITIVITY VERNIER Control . . full CCW
METER RANGE Selector . . . . . . . . . . 0 db
FREQUENCY RANGE Selector . . . . . . . Xl
Frequency Dial . . . . . . . . . . . . . . .5
Set oscillator for approximately 1 volt output at
5 cycles.
d. Set filter box for 5 cycles.
e. Adjust oscillator amplitude for an indication of
+2 db on the Distortion Analyzer meter.
f. Switch Distortion Analyzer FUNCTION selector
to DISTORTION. Adjust frequency dial and BALANCE
controls for a null meter indication.
(If reading is in
lower 1/3 of meter scale, decrease METER RANGE
selector setting. )
g. The meter indication at “null” shall be at least
-8 db on the -60 db METER RANGE which is equiva-
lent to -70 db. Note reading.
Figure 5-2. Instrument Induced Distorition and Automatic Control Loop Test Setup
5-2

Table of Contents

Other manuals for HP 333A

Related product manuals