EasyManua.ls Logo

HP 400D - Page 10

HP 400D
29 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section
II
Paragraphs
3-22
to
3-25
+
2
(meter
indication)
+30
(RANGE
switch
position)
+32
(sum)
+
8
(correction
factor
from
the
Impedance
+40
dbm
Correction
Graph)
Example
2
For
the
same
conditions
as
given
above,
except
that
the
measurement
is
made
across
an
impedance
of
60,000
ohms,
the
level
in
dbm
is
obtained
as
follows:
+
2
(meter
indication)
+30
(RANGE
switch
position)
+32
(sum)
~-20
(Correction
factor
from
the
Impedance
+12
dbm
Correction
Graph)
3-22.
USE
OF
VOLTMETER
AMPLIFIER.
3-23.
The
amplifier
in
the
voltmeter
may
be
used
for
amplifying
weak
signals.
With
full~scale
meter
deflec-
tion,
the
open-circuit
output
of
the
amplifier
is
approxi-
mately
0.15
volt
rms
regardless
of
the
RANGE
switch
position.
The
impedance
looking
into
the
OUTPUT
terminals
is
approximately
50
ohms.
The
frequency
3-4
T.O.
33A1-12-349-1
response
and
calibration
of
the
voltmeter
may
be
affected
by
the
impedance
of
a
load
applied
to
the
OUTPUT
terminals.
To
check
the
effect
of
the
applied
load:
observe
the
meter
reading
obtained
with
no
load
connected
to
the
OUTPUT
terminals
and
then
note
any
shift
of
reading
when
the
external
circuit
is
connected
to
the
OUTPUT
terminals.
If
the
shift
is
negligible,
the
measurement
is
not
being
affected
appreciably
by
the
load.
Whenever
the
input
signal
is
changed,
i.e.,
a
different
frequency
or
band
of
frequencies
is
applied,
repeat
the
quick
check
described
above.
3-24.
Maximum
gain
from
the
amplifier
is
obtainable
only
on
the
lowest
(.001
volts)
range,
since
output
level
is
the
same
for
all
bands.
This
is
due
to
the
10-db
amplification
loss
per
step
inserted
by
the
RANGE
switch
as
it
is
turned
clockwise.
Amplification
may
also
be
obtained
on
the
.003,
.01,
.03,
and
1
volt
ranges.
3-25.
When
the
voltmeter
is
used
as
an
amplifier,
select
a
range
which
gives
a
meter
deflection
near
full
scale.
Off-scale
signals
more
than
twice
the
value
of
the
position
of
the
RANGE
switch
will
cause
severe
distortion.
00102
-2

Other manuals for HP 400D

Related product manuals