EasyManua.ls Logo

HP 400D - SECTION Ill; Instrument Turn-On; General Operating Information

HP 400D
29 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
T.O.
33A1-12-349-1
Section
III
Paragraphs
3-1
to
3-9
SECTION
Ill
OPERATING
INSTRUCTIONS
3-1.
INSTRUMENT
TURN-ON.
3-2.
The
voltmeter
is
ready
for
use
as
received
from
the
factory
and
will
give
specified
performance
after
a
few
minutes
warmup.
See
Section
II
for
information
regarding
connection
to
the
power
source
and
to
the
voltage
to
be
measured.
Controls
are
shown
in
figure
3-1.
3-3.
GENERAL
OPERATING
INFORMATION.
3-4.
METER
ZERO
CHARACTERISTIC.
When
the
Model
400D
and
400H
Voltmeters
are
turned
off,
the
meter
pointer
should
rest
exactly
on
the
zero
calibration
mark
on
the
meter
scale.
If it
does
not,
zero-set
the
meter
as
instructed
in
paragraph
5-7.
The
meter
supplied
in
the
Model
400L
Voltmeter
is
not
provided
with
a
mechanical
meter
zero
adjustment.
When
the
voltmeter
is
turned
on
with
the
INPUT
terminals
shorted,
the
meter
pointer
may
deflect
upscale
slightly;
this
deflection
does
not
affect
the
accuracy
of
a
reading.
NOTE
When
the
voltmeter
RANGE
switch
is
set
to
the
lowest
ranges
and
the
INPUT
terminals
are
not
terminated
or
shielded,
noise
pickup
can
be
enough
to
produce
up
to
full-scale
meter
deflec-
tion.
This
condition
is
normal
and
is
caused
by
stray
voltages
in
the
vicinity
of
the
instru-
ment.
For
maximumaccuracy
on
the
-001-volt
range,
the
voltage
under
measurement
should
be
applied
to
the
voltmeter
through
a
shielded
test
lead.
3-5.
METER
SCALES.
The
two
voltage
scales
on
each
of
the
voltmeter
models
are
related
to
each
other
by
a
factor
of
1710
(10
db).
In
conjunction
with
the
calib-
rated
RANGE
switch
steps,
this
provides
an
intermediate
range
step
spaced
10
db
between
"power
of
ten"
ranges,
which
are
20
db
apart.
The
relationship
of
the
DECIBELS
scale
to
the
0
to
1
VOLT
scale
is
determined
by
making
0
db
on
the
DECIBELS
scale
equal
to
the
voltage
required
to
produce
1
milliwatt
in
600
ohms
(0.775
volts).
Thus,
the
DECIBELS
scale
reads
directly
in
dbm
(decibels
referred
to
one
milliwatt)
across
a
600-ohm
circuit,
and
can
be
used
to
measure
absolute
level
of
sine
wave
signals.
It
can
also
be
used
to
measure
relative
levels
of
any
group
of
signals
which
have
the
same
waveform,
across
any
constant
circuit
impedance.
The
RANGE
switch
changes
voltmeter
sensitivity
in
10-db
steps
accurate
to
within
+1/8
db.
The
RANGE
switch
position
indicates
the
value
of
a
full-scale
meter
reading.
3-6.
CONNECTIONS.
Voltmeter
test
leads
must
be
provided
by
the
user.
The
type
of
leads
and
probes
used
will
depend
upon
the
application,
as
listed
below:
a
For
connection
to
low-impedance
signal
sources,
plain
wire
leads
often
are
sufficient.
00102-2
b.
For
high-impedance
sources,
or
where
noise
pickup
is
a
problem,
low-capacity
shielded
wire
must
be
used
with
a
shielded,
dual
banana
plug
for
connection
to
the
voltmeter
terminals.
c.
If
a
probe
is
used,
it
should
also
be
shielded
to
prevent
pickup
from
the
hand.
d.
For
signals
above
a
few
hundred
kilocycles,
the
capacity
of
the
test
leads
must
be
kept
to
a
minimum
by
using
very
short
leads,
preferably
unshielded.
An
alligator
clip
should
be
used
at
the
test
end
so
that
connection
can
be
made
without
adding
the
capacity
of
the
user's
hands.
3-7.
MAXIMUM
INPUT
VOLTAGE,
Do
not
apply
more
than
600
volts
dc
to
the
INPUT
terminals.
To
do
so
ex~-
ceeds
the
voltage
rating
of
the
input
capacitor.
3-8.
If
an
applied
voltage
momentarily
exceeds
the
selected
full-scale
voltmeter
sensitivity,
a
few
seconds
may
be
required
for
circuit
recovery,
but
no
damage
will
result.
3-9.
INPUT
VOLTAGE
WAVEFORM.
The
voltmeter
is
calibrated
to
indicate
the
root-mean-square
value
of
a
sine
wave;
however,
meter
pointer
deflection
is
proportional
to
the
average
value
of
whatever
waveform
is
applied
to
the
input.
If
the
input
signal
waveform
is
not
a
sine
wave,
the
reading
will
be
in
error
by
an
amount
dependent
upon
the
amount
and
phase
of
the
harmonics
present,
as
shown
in
figure
3-2
below.
When
harmonic
distortion
is
less
than
about
10%,
the
error
which
results
is
negligible.
INPUT
VOLTAGE
CHARACTERISTICS
METER
INDICATION
Fundamental
=
100
100
Fundamental
+10%
100
2nd
harmonic
Fundamental
+20%
2nd
harmonic
100-102
Fundamental
+50%
100-110
2nd
harmonic
Fundamental
+10%
3rd
harmonic
Fundamental
+20%
3rd
harmonic
Fundamental
+50%
3rd
harmonic
Note:
This
chart
is
universal
in
application
since
these
errors
are
inherent
in
all
average-respond-
ing
type
voltage-measuring
instruments.
96-104
94-108
90-116
Figure
3-2.
Effect
of
Harmonics
on
Voltage
Measurements
3-1

Other manuals for HP 400D

Related product manuals