EasyManua.ls Logo

HP 400D - Set the RANGE

HP 400D
29 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section
TI
Paragraphs
3-10
to
3-16
=—
POSSIBLE
GROUND
LOOP
+
T.O.
33A1-12-349-1
INSERT
UNGROUNDED
ADAPTER
TO
BREAK
o
T
USE
GROUND
LOOP
eae
ONE
SIDE
OF
POWER
LINE
IS
ALSO
GROUNDED
Figure
3-3.
Test
Setup
for
Avoiding
Ground
Loop
3-10. Since
the
voltmeter
meter
deflection
is
propor-
tional
to
the
average
value
of
the
input
waveform,
it
is
not
adversely
affected
by
moderate
levels
of
random
noise.
The
effect
that
noise
has
on
the
accuracy
of
the
meter
reading
depends
upon
the
waveform
of
the
noise
and
upon
the
signal-to-noise
ratio.
A
square
wave
has
the
greatest
effect,
a
sine
wave
intermediate
effect,
and
"white"
noise
has
the
least
effect
on
the
meter
reading.
3-11.
If
the
noise
signal
is
a
50%
duty
cycle
square
wave
and
the
signal-to-noise
ratio
is
10:1
(between
peak
voltages),
the
error
will
be
about
1%
of
the
meter
reading.
If
the
noise
signal
is
"white"
noise
and
the
signal-to-noise
ratio
10:1,
the
error
is
negligible.
3-12.
LOW-LEVEL
MEASUREMENTS
AND
GROUND
CURRENTS.
3-13.
When
the
voltmeter
is
used
to
measure
signal
levels
below
a
few
millivolts,
ground
currents
in
the
meter
test
leads
can
cause
an
error
in
meter
reading.
Such
currents
are
created
when
two
or
more
ground
connections
are
made
between
the
instruments
of
a
test
setup
and/or
between
the
instruments
and
the
power
line
ground.
Two
ground
connections
complete
an
electrical
circuit
(ground
loop)
for the
voltages
which
are
generated
across
all
instrument
chassis
by
stray
fields,
particularly
the
fields
of
transformers.
These
ground
currents
can
be
minimized
by
disconnecting
the
ground
lead
in
the
power
cord
from
either
the
voltmeter
or
the
signal
source
being
measured,
at
the
power
outlet
as
shown
in
figure
3-3,
and
by
making
sure
that
in
the
test
setup
no
other
ground
loop
is
formed
that
can
cause
a
ground
current
to
flow
in
the
voltmeter
test
leads.
Although
the
resultant
voltage
developed
across
a
test
lead
is
in
the
order
of
micro-
volis,
it
is
enough
to
cause
noticeable
errors
in
measurements
of
a
few
millivolts.
The
presence
of
ground
currents
can
sometimes
be
determined
by
simply
changing
the
grounds
for
the
instruments
in
the
3-2
setup
and
watching
for
a
change
in
meter
reading.
If
changing
the
ground
system
causes
a
change
in
meter
reading,
ground
currents
are
present.
3-14.
MEASUREMENT
OF
VOLTAGE.
3-15.
The
meter
has
two
VOLTS
scales,
0
to
1
and
0
to
3.
When
the
RANGE
switch
is
set
to
.001,
.01,
-1,
1,
10,
or
100
VOLTS,
read
the
0
to
1
scale.
When
the
RANGE
switch
is
set
to
.003,
.03,
.3,
3,
30,
or
300
VOLTS,
read
the
0
to
3
scale.
The
lower
(black)
signal
INPUT
and
OUT-
PUT
terminals
and
the
instrument
case
are
connected
tothe
power
system
ground
when
the
instrument
is
used
witha
standard
three-
terminal
(grounding)
receptacle.
Connect
only
ground-potential
circuits
to
the
black
INPUT
and
OUTPUT
terminals.
3-16.
Operate
the
instrument
as
follows:
a.
Connect
the
voltmeter
to
the
a-c
power
source.
b.
Turn
the
Power
switch
ON
and
allow
a
warmup
period
of
approximately
five
minutes.
c.
Disconnect
any
external
equipment
from
the
OUT-
PUT
terminals.
d.
Set
the
RANGE
switch
to
the
VOLTS
range
which
will
read
the
voltage
to
be
measured
at
mid-scale
or
above.
If
in
doubt,
select
a
higher
VOLTS
range.
e.
Connect
the
voltage
to
be
measured
to
the
INPUT
terminals.
00102-2

Other manuals for HP 400D

Related product manuals