L 
•  Page 5-2, Paragraph 5-13, 4328A PERFORMANCE CHECKS: 
Add the 
following NOTE to paragraph 5-14. 
NOTE 
Before starting the  Performance Check for the 4328A Option 001 , confirm 
that 
the 
battery  has enough charge  by  pressing  the  BATT. 
TEST  button.  If the  pointer 
of 
meter is  out 
of 
the 
BAT. 
CHARGED range on  the meter scale,  recharge the battery 
following  the  instructions given  in  Paragraph  3-9.  The  4328A Option 
001 
will 
not 
operate properly, if the battery does not have enough charge. 
CHANGE 1 
..,. 
Section 6,  Replaceable Parts, Table 6-2, Reference Designation Index: 
See 
Table 1, Parts Information. 
CHANGE 2 
Section 6, Replaceable Parts, Table 6-2, Reference Designation Index: 
See Table 
1, Parts Information. 
Page 4-7, Figure 4-5 and Page 8-19, Figure 8-17, 
A2 
PHASE-SENSITIVE AMPLIFIER 
Assy.: 
Change the value 
of 
R230" to 56k0. 
Page 8-21, Figure 8-19: 
Change the value 
of 
R3 
to 33k0. 
Page 8-23, Figure 8-21: 
Change the value 
of 
R3 
to 33k0. 
CHANGE 3 
Page 6-8, Table 6-2, Reference Designation Index: 
See 
Table 1, Parts Information. 
Page 8-17, Figure 8-14: 
A3 
OSCILLATOR Assy. 
Change the value 
of 
R335 to 133kn. 
-
2-
---
---~~-~--
---··~-~----------------
' 
•