Model 436A
General Information
Table
1-2.
Recommended
Test
Equipment
Instrument
Type
Critical
Specifications
Suggested
Use*
Model
Range Calibrator
Chopped
de
output
for each range referenced to 1
mW
range
HP
11683A
P,A,T
Digital Voltmeter
Function:
DC,
resistance
HP
3456A
P,A,T
Range Resistance:
200
ohms
Vdc: 100
mVdc,
1000
mVdc,
10 Vdc, lOOVdc
10Mr2
input impedance
6-digit resolution
(±0.05%
of
reading,
±0.02%
of
range)
Power Meter
Range: 1
mW
HP
432A
P,A
Transfer Accuracy (input -to-output):
0.2%
Thermistor
SWR:
1.05, 50
MHz
HP
478A-H75**
P,A
Mount
Accuracy:
±0.5%
at
50
MHz
or
HP
478A-H76**
Counter
Frequency Range:
220
Hz, 50
MHz
HP
5315A
A
Sensitivity:
100
m
Vrms
Accuracy:
0.01%
Oscilloscope
Bandwidth:
de
to
50
MHz
HP
180C/ T
Vertical Sensitivity: 0.2V/division
1801A/1821A
Horizontal Sensitivity: 1 ms/division
Logic Analyzer
Clock Input:
60
kHz
HP
1600A
T
Trigger Word: 8 Bits
Bit Input: TTL
Display Word: 8 Bits
•p
=Performance
Tests; A
=Adjustments;
T =Troubleshooting
**For maximum accuracy
the
478A-H75
should
be calibrated
by
the
National Bureau of
Standards.
The
478A-H76, which includes
HP
Standards
Laboratory calibration,
may
be used with a measurement
system accuracy of
0.58%
instead
of
0.5%.
1-5/1-6