Error Messages
This section lists the error messages that are displayed on the analyzer display or transmitted
by the instrument over HP-IB. Each error message is accompanied by an explanation, and
suggestions are provided to help in solving the problem. Where applicable, references are
provided to the related chapter of the appropriate manual.
When displayed, error messages are preceded with the word \
CAUTION:
." That part of the
error message has been omitted here for the sake or brevity. Some messages without the
\
CAUTION:
" are for information only, and do not indicate an error condition. The messages are
listed rst in alphabetical order because the displayed messages do not contain the message
number. The messages are then listed in numerical order to make them easier to nd if they
are read over the HP-IB.
In addition to error messages, The analyzer's status is indicated by status notations in the left
margin of the display. Examples are
3
,
Cor
, and
Avg
. Sometimes these appear together with
error messages.
Error Messages in Alphabetical Order
A
10
ADDITIONAL STANDARDS NEEDED
Error correction for the selected calibration class cannot be computed until all the necessary
standards have been measured.
84
ANALYZER TYPE MISMATCH
The analyzer receives a command that is not available for the current analyzer type. Please
conrm HP-IB command or change analyzer type before sending the command.
B
17
BACKUP DATA LOST
Data checksum error on the battery backup memory has occurred. The battery is recharged for
approximately 10 minutes after power was turned
on
.
0
160
Block data error
This error, as well as errors
0
161 and
0
168, are generated when analyzing the syntax of a
block data element. This particular error message is used if the analyzer cannot detect a more
specic error.