Error Messages in Alphabetical Order
1
CAN'T SET RBWAUTO IN ZERO SPAN
The RBW AUTO mode cannot be selected in the zero span. The RBW must be specied
manually in the zero span. (spectrum analyzer mode only).
127
CAN'T SET SWEEP TIME AUTO IN ZERO SPAN
The automatic sweep time cannot be in zero span of the spectrum analyzer mode. (The
network analyzer mode allows that the automatic sweep time is turned on.)
0
281
Cannot create program
Indicates that an attempt to create a program was unsuccessful. A reason for the failure might
include not enough memory.
0
140
Character data error
This error, as well as errors
0
141 through
0
148, are generated when analyzing the syntax of a
character data element. This particular error message is used if the analyzer cannot detect a
more specic error.
0
148
Character data not allowed
A legal character data element was encountered where prohibited by the analyzer
.
0
144
Character data too long
The character data element contains more than twelve characters (see IEEE 488.2, 7.7.1.4).
0
100
Command error
This is a generic syntax error that the analyzer cannot detect more specic errors
. This code
indicates only that a command error, as dened in IEEE 488.2, 11.5.1.1.4, has occurred.
0
110
Command header error
An error was detected in the header. This error message is used when the analyzer cannot
detect the more specic errors described for errors
0
111 through
0
119.
75
COMMAND IGNORED - SEGMENT NOT DONE YET
The HP-IB command the analyzer received is ignored, because the segment is editing (
HP-IB
only
). Send
LIMSDON
(limit segment done) or
SDON
(segment done) to terminate editing
segment. (See
Programming Manual
.)
269
COMPENSATION ABORTED
Compensation data acquisition process is aborted.
267
COMPENSTATION REQUIRED
Compensation is required. Perform compansation to obtain compensation data.
Messages-3