Error Messages in Alphabetical Order
143
FLOATING POINT ERROR OCCURED
Indicate that a oating point error occured in the analyzer. Data processing may not be correct.
This error message is used when an internal application was executed for illegal data sent from
an external device, or when an internal software bug was detected. Contact your nearest
Hewlett-Packard oce.
83
FORMAT NOT VALID FOR MEASUREMENT
The conversion function except the 1/S and the multiple phase modes is not valid for the
Smith, admittance, and SWR formats.
131
FREQUENCY SWEEP ONLY
The sweep type must be frequency sweep when the center step size is set.
G
0
105
GET not allowed
A Group Execute Trigger (GET) was received within a program message (see IEEE 488.2, 7.7).
H
0
240
Hardware error
Indicates that a legal program command or query could not be executed because of a hardware
problem in the analyzer. Denition of what constitutes a hard ware problem is completely
device-specic. This error message is used when the analyzer cannot detect the more specic
errors described for errors
0
241 through
0
249.
0
241
Hardware missing
A legal program command or query could not be executed because of missing analyzer
hardware.For example, an option was not installed.
0
111
Header separator error
A character that is not a legal header separator was encountered while parsing the header.For
example, no white space followed the header, thus
*SRE4
is an error.
0
114
Header Sux out of range
The value of a numeric sux attached to a program mnemonic makes the header invalid.
I
0
224
Illegal parameter value
Used where exact value, from a list of possibilities, was expected.
Messages-6