Error Messages in Numerical Order
Error Messages in Numerical Order
0 - 100
+0
No error
The error queue is empty. Every error in the queue has been read (
OUTPERRO?
query) or the
queue was cleared by power-on or the
3
CLS
command.
1
CAN'T SET RBWAUTO IN ZERO SPAN
The RBW AUTO mode cannot be selected in the zero span. The RBW must be specied
manually in the zero span. (spectrum analyzer mode only).
10
ADDITIONAL STANDARDS NEEDED
Error correction for the selected calibration class cannot be computed until all the necessary
standards have been measured.
11
CALIBRATION REQUIRED
No valid calibration coecients were found when you attempted to turn calibration
on
.
12
NO CALIBRATION CURRENTLY IN PROGRESS
The
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
RESUME CAL SEQUENCE
softkey is not valid unless a calibration is in progress
. Start a new
calibration.
13
CALIBRATION ABORTED
The calibration in progress was terminated due to a change of the active channel or stimulus
parameters.
14
NOT VALID FOR PRESENT TEST SET
The calibration requested is inconsistent with the test set present. This message occurs in the
following situations:
A full 2-port calibration is requested with a test set other than an S-parameter test set.
A one-path 2-port calibration is requested with an S-parameter test set (this procedure is
typically used with a transmission/reection test set).
15
EXCEEDED 7 STANDARDS PER CLASS
A maximum of seven standards can be dened for any class.
16
CURRENT PARAMETER NOT IN CAL SET
HP-IB only.
Correction is not valid for the selected measurement parameter.
17
BACKUP DATA LOST
Data checksum error on the battery backup memory has occurred. The battery is recharged for
approximately 10 minutes after power was turned
on
.
Messages-15