Table 2-1. Source Level Flatness Test Settings 1
HP 4395A
Center Frequency
1 MHz
3 MHz
10 MHz
45 MHz
75 MHz
110 MHz
140 MHz
190 MHz
235 MHz
290 MHz
345 MHz
400 MHz
445 MHz
470 MHz
500 MHz
c. Wait for the power meter reading to settle. Then record the reading in the calculation
sheet (\Power Meter Reading" column).
d. Change the HP 4395A center frequency in accordance with T
able 2-1 and repeat step 5-c
for each frequency.
e. Calculate test results using the equation given in the calculation sheet. Record the test
results in the performance test record.
6.
Source Level Flatness Test (Low Frequencies)
a. Connect the test equipment as shown in Figure 2-3.
Figure 2-3. Source Level Flatness Test Setup
b. Initialize the multimeter. Then set the controls as follows:
2-6 Performance Tests