Note
Connect the signal generator's 10 MHz frequency reference output to the
HP 4395A EXT REF Input on the rear panel as shown in Figure 2-24. With this
conguration, both the signal generator and the HP 4395A are phase locked to
the same reference frequency to obtain a stable measurement.
Figure 2-24. Frequency Response Test Setup 1
c. Initialize the signal generator. Then set the amplitude to
0
14 dBm.
d.
Press
4
Meas
5
,
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ANALYZER TYPE
,
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SPECTRUM ANALYZER
,
4
Preset
5
to initialize the HP 4395A.
Then set the controls as follows:
Control Settings Key Strokes
Frequency Span: 1 kHz
4
Span
5
,
4
1
5
,
4
k/m
5
Reference Level:
0
10 dBm
4
Scale Ref
5
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
REFERENCE VALUE
,
4
-
5
,
4
1
5
,
4
0
5
,
4
x1
5
Scale/Division: 1 dB/Div
4
Scale Ref
5
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SCALE/DIV
,
4
1
5
,
4
x1
5
e.
Press
4
Meas
5
,
NNNNN
R
to set the HP 4395A to the R input.
f.
Press
4
Scale Ref
5
,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ATTENUATOR MENU
,
NNNNNNNNNNNNNNNNNNNNNNN
ATTEN R
,
4
1
5
,
4
0
5
,
4
x1
5
to set the input attenuator to
10 dB.
g. Press
4
Bw/Avg
5
,
NNNNNNNNNNNNNNNNNNNN
RES BW
,
4
3
5
,
4
0
5
,
4
0
5
,
4
k/m
5
in the rst column of Table 2-22. Table 2-22 lists
the test settings at frequencies
1 MHz.
h. On the signal generator, set the frequency to the rst frequency 50 MHz in the rst
column of Table 2-22.
2-62 Performance Tests