Operation
and
Programming
3-222.
Power/Warm-Up
3-223.
The
HP
Model
5335A
requires
a
power
source
of
100,
120,
220,
or
240V
ac,
4-5%,
-10%,
48
to
66
Hz
single
phase.
The
selection
of
line
voltage
and
input
power
fuse
is
described
in
Section
II,
paragraph
2-5,
Preparation
for
Use.
3-224.
The
5335A
has
a
two-position
power
switch,
STBY
and
ON.
For
5335A
Option
010
only,
it
is
important
that
the
instrument
remain
connected
to
the
power
source
in
the
STBY
mode
when
not
in
use.
This
supplies
power
to
the
crystal
oven
maintaining
a
constant
oven
tempera¬
ture,
thus
eliminating
the
need
for
a
warm-up
period.
When
the
STBY
mode
is
not
used
and
power
is
disconnected
from
the
instrument,
allow
30
minutes
from
the
application
of
external
power
in
the
ON
mode
for
the
instrument
(crystal
oven)
to
warm-up.
POWER
IS
ALWAYS
PRESENT
AT
THE
LINE
SWITCH
AND
TRANSFORMER,
AND
UNREGULATED
DC
IS
PRESENT
WHEN¬
EVER
THE
LINE
CORD
IS
ATTACHED.
UNPLUGGING
THE
POWER
CORD
IS
NECESSARY
TO
REMOVE
ALL
POWER
FROM
THE
INSTRUMENT.
3-225
OPERATOR'S
CHECK
3-226.
The
following
procedures
will
verify
the
basic
operation
of
the
HP5335A
Universal
Counter.
These
tests
are
not
intended
to
verify
the
overall
accuracy
or
performance
specifications
of
the
instrument.
They
should,
however,
provide
the
operator
a
quick
method
of
determining
that
the
counter
is
operating
properly.
The
tests
are
provided
in
two
levels;
a
DISPLAY
CHECK
and
a
FRONT
PANEL
CHECK.
The
operator
should
perform
both
tests.
A.
DISPLAY
CHECK
To
perform
the
HP5335A
DISPLAY
CHECK,
momentarily
press
the
key
labeled
“CHECK".
Verify
that
all
display
annunciators,
except
“STBY"
and
the
A
and
B
Channel
Trigger
lights,
cycle
ON
and
OFF.
Pressing
any
Function
key
will
return
normal
operation.
3-57
Model
5335A